Before treating different voltage readings as a fault, compare the same quantity, coupling and operating condition. DC average, RMS, peak and peak-to-peak are different measurements.
First identify the numbers being compared
FNIRSI lists multimeter functions and scope readings including average, RMS and peak-to-peak for the 2C53P. A multimeter DC voltage value is not automatically comparable with the tallest point of a scope trace. Write down the selected mode and units on both screens. The scope’s reported RMS may include a DC component or use the displayed acquisition window; confirm the exact definition for the instrument rather than assuming it matches an AC meter mode.
Use a stable low-voltage source for the first comparison
Begin with a suitable steady low-voltage source and no changing load. Connect voltage inputs according to the exact model instructions and never place a current input directly across the source. Confirm a safe common reference. If switching modes requires moving leads or stopping acquisition, do that as instructed; a combined instrument’s modes need not measure the same instant. Record the source behaviour during each reading.
Check coupling and probe scale
A scope in AC coupling removes the steady component, so a healthy DC rail can appear centred around zero. A physical 10X probe with the channel still set to 1X can produce a factor-of-ten display error. Check both settings before evaluating accuracy. Use DC coupling when comparing the rail’s steady level; choose sufficient vertical range so the waveform is not clipped by the displayed acquisition.
A numerical example without a hardware claim
For an ideal sine wave centred on zero with a 1 V peak, peak-to-peak is 2 V and RMS is about 0.707 V. Its average over complete cycles is zero. Those different numbers can all describe the same signal. Add a DC offset and the relationships change again. Real meter AC bandwidth, waveform response and scope sampling can impose additional limitations, so do not use this example as a specification for the 2C53P.
Decide what an unresolved difference proves
Repeat under the same source condition and compare the relevant manufacturer specifications and stated uncertainty. A mode-to-mode disagreement can arise from configuration or source variation and does not identify a failed channel by itself. If a calibrated value is needed, compare with an appropriate independent reference and method. The purchase decision should reflect which functions you need, not an expectation that every displayed voltage must be identical.
Instrument and next step
This is a proposed low-voltage workflow, not a hardware test or a measured fault report. Check the exact instrument, probe and circuit documentation before connecting. If your current equipment already resolves the question, keep using it.