74HC123 One-Shot Pulse Looks Too Long: Check Retriggering
DigitalOsc editorial team · Updated 20 November 2026
A retriggerable one-shot can extend its output when another valid trigger arrives. Compare trigger and output before changing the timing resistor or capacitor.
Check the exact monostable variant
The TI CD74HC123 documentation describes dual retriggerable monostable multivibrators with reset. Confirm the complete marking and data sheet because a 123-family suffix, logic family or different one-shot can change timing and permitted inputs. Identify which trigger input is active in the actual circuit and how the other input and reset are held. An unconnected control input can invalidate a tidy-looking timing calculation.
Inspect the input stream before the timing parts
Begin with a known safe low-voltage source and a repetition interval long enough for one output pulse to finish. Use a compatible reference, check the logic levels and trigger polarity, and avoid changing component values while the circuit is powered. This baseline separates a single trigger from a burst of edges produced by the actual source or its noise.
Record trigger and output in one window
Place one input on the active trigger path and the other on the selected Q output. Choose a timebase that includes the full output pulse and the next incoming edge. If a valid second trigger arrives before the first interval ends, the output may remain active longer in accordance with the retrigger function. A long output is therefore not sufficient evidence that the capacitor is wrong.
Use a controlled comparison
Record one isolated trigger, then a controlled pair or burst within the allowed input conditions. Compare pulse duration and trigger spacing with the exact part’s truth table and timing guidance. For example, a nominal 10 ms design observed under 5 ms trigger spacing is a different experiment from one triggered once per second. Those numbers illustrate a comparison; they are not component values or a measured performance claim.
Only then investigate the RC timing path
If the isolated-trigger result remains wrong, check the actual resistor, capacitor, leakage, reset and wiring against the manufacturer’s formula and operating conditions. Do not transplant a timing equation from an NE555 guide. Two inputs help show whether extension coincided with another trigger; neither retail model is presented as a calibrated pulse-width reference. Retain the capture and circuit conditions before deciding which part needs replacement.
Instrument and next step
This is a proposed low-voltage workflow, not a hardware test or a measured fault report. Check the exact instrument, probe and circuit documentation before connecting. If your current equipment already resolves the question, keep using it.
DigitalOsc links to Payverza as the retailer. Current price, availability and delivery are shown there; a product link does not certify accessories, isolation or the suitability of a particular circuit.
Sources and scope
- HC123 official product or application documentation
- Pico Technology: choosing and using an oscilloscope
Sources checked 20 November 2026. Worked numbers are illustrative; they are not measurements of either retail instrument.